2 Tlab PRP tubes with anticoagulant

10 ml vacuum tubes for preparing autologous PRP in a single centrifugation. Each tube contains 1 ml of sodium citrate, with no gel and no Ficoll. Gamma-sterilised, non-haemolytic, pyrogen-free and single use. CE Class IIb medical device. Dimensions: 105 mm long, 15 mm diameter, 18 mm cap. Box of 2 tubes.

T-1003
€32.00
Tax excluded Restocking in progress. Estimated delivery between 5 and 10 days.

Volume discounts

€0.83
Buy 10 or more
Save: €8.33
€1.67
Buy 50 or more
Save: €83.33
Replenishment in progress
Quantity

  CE Certification
  Delivery all over Europe
  Free shipping for orders over 500€ HT
Description

Two 10 ml vacuum tubes, prefilled with 1 ml of sodium citrate, to prepare pure PRP in a single centrifugation.

Box of 2 T-LAB PRP tubes with anticoagulant

T-LAB kit, reference T-1003

Specifications

Composition

  • Volume: 10 ml, vacuum tube
  • Anticoagulant: 1 ml of sodium citrate
  • No separator gel, no Ficoll
  • Non-haemolytic, pyrogen-free
  • Gamma sterilisation, single use
  • CE Class IIb medical device

Dimensions and packaging

  • Length: 105 mm
  • Tube diameter: 15 mm
  • Cap diameter: 18 mm
  • Box of 2 tubes (reference T-1003)

Centrifugation protocol

Pure PRP. The T-LAB protocol is 830 g (RCF) for 4 minutes, in a single centrifugation.

Set RCF rather than RPM. The speed needed to reach 830 g depends on the rotor radius, so it varies from one centrifuge to another. If your device only displays RPM, convert using the rotor radius given in its manual.

Balancing. Place the two tubes from the box opposite each other in the rotor.

PRP preparation with T-LAB tubes

Instruction video

Test report results

T-LAB in vitro performance report (RPR.PTN.06, 2019), measured on PRP prepared with two tubes:

  • Platelet concentration: 2.40 times that of whole blood
  • Platelet recovery: 80.1%
  • Platelet purity: 95.6%
  • PRP volume obtained: 5.4 ml from two tubes

Documents

picture_as_pdfTest reportPDF

In vitro performance, RPR.PTN.06 (2019)

picture_as_pdfSafety data sheetPDF

SDS for T-LAB PRP tubes

Frequently asked questions

Are these tubes compatible with my centrifuge?expand_more
The tube is 105 mm long and 15 mm in diameter, with an 18 mm cap. Check in your centrifuge manual that the tube holder accepts these dimensions, without the reducer used for 13 mm tubes.
Which setting should I use if my centrifuge only displays RPM?expand_more
Convert 830 g to RPM using the rotor radius, with the formula RCF = 1.118 × 10-5 × radius (cm) × RPM². The radius is given in the device manual.